How to Read a Peptide COA

A certificate of analysis (COA) is the laboratory document that reports the test results for one specific lot. The lot number on the vial should match the lot number on the COA. This guide explains the parts of a typical peptide COA, the tests a laboratory may run, what a complete result looks like, and warning signs that a document is incomplete.

Not every COA includes every test described here. Read the actual report for the lot in hand, and use the lot number or the QR code on the label to find the matching file on our COA / Lab Results page.

Laboratory reference only This page is general reference for laboratory researchers. All USAPI products are for research use only and are not for human or veterinary use. It is not a certificate of analysis and is not a substitute for the lot-specific COA.

Anatomy of a COA

A complete report usually includes the following parts. The numbered markers match the example figure below.

1

Laboratory name

The testing laboratory’s name and contact details, and often a unique report or sample number. A complete COA makes the lab identifiable. USAPI lots are tested by an independent laboratory before release.

2

Product and lot number

The compound name, and often a catalog number or sequence. The lot number must match the code printed on the vial label — and the label’s QR code opens the matching file on the COA / Lab Results page.

3

Dates

When the sample was received, tested and reported. Some reports also list a retest date for the lot. A report with no dates is incomplete.

4

Method

The analytical method used for each test, such as reverse-phase HPLC at a stated wavelength, or LC-MS. A result without a method is harder to interpret.

5

Specification vs. result

A specification is the acceptance criterion (for example, HPLC purity ≥99%). The result is what was measured for that lot. Both should appear, with units.

6

Signatures

The analyst and a reviewer or approver, with dates. An unsigned report is a weak document.

7

Supporting figures

For purity and identity, a chromatogram and a mass spectrum (or a peak list) should be attached or embedded. A number without the figure is only half a result.

Example COA (illustrative only)

The values in this figure are placeholders (99.X %, XXXX.X Da, EXAMPLE-0000). They are not results for any real lot.

  • Laboratory name. Who performed the tests, and a report ID.
  • Lot number. Must match the vial label.
  • Dates. When the lot was tested and the report was issued.
  • Method. How each result was measured.
  • Specification vs. result. The limit and the measured value, with units.
  • Signatures. Analyst and reviewer, with dates.
  • Supporting figures. Chromatogram and mass spectrum (or a peak list).

Tests at a glance

Use this table to scan a report, then read the section for any test that appears on the COA in front of you.

TestWhat it tells youTypical methodWhat to look for
HPLC purityShare of the main peptide versus related impuritiesReverse-phase HPLC (UV)A number at or above the stated spec (USAPI: ≥99%), plus a chromatogram
Identity (MS)Whether the material is the labeled compoundLC-MS or MALDI-TOFExpected and observed mass both listed and in agreement
Net peptide contentPeptide mass versus total powder (excludes salt and water)Amino acid analysis or nitrogen analysisA numeric content with the method named; not the same as HPLC purity
Water contentResidual moisture in a lyophilized solidKarl Fischer titrationA numeric percentage within the specification on that COA
EndotoxinBacterial endotoxin levelLAL assay (or a recombinant equivalent)A numeric result in endotoxin units, versus a stated specification
SterilityWhether viable microorganisms were detectedCulture in nutrient media for a defined periodMethod, laboratory and a clear result; many research COAs do not include this test
Heavy metals / elemental impuritiesTrace elemental impurities (for example Pb, As, Cd, Hg)Typically ICP-MSNamed elements, each with a result, units and a limit
Residual solventsTrace solvents left from synthesis or purificationGas chromatographyEach solvent named, with a result and a limit
TFA / counterion contentAmount of salt form (TFA or acetate) in the powderIon chromatography or 19F NMRSalt form identified and, when reported, quantified
AppearanceVisual description of the lyophilized cake or powderVisual inspectionA description that matches a dry solid, with any unexpected color noted
SolubilityWhether the lot dissolved in a named laboratory solventObservation under stated conditionsSolvent, conditions and observation all stated; not a use instruction
Bioburden / microbial limitsCount of viable microorganisms (not a sterility test)Plate counts on appropriate mediaNumeric counts, units, limits and the method
Amino acid analysisResidue composition versus the expected sequenceHydrolysis, then chromatographyFound versus theoretical residues, with method and notes on labile residues

The tests, one by one

Each section covers what the test measures, how it is typically done, what a complete result looks like, and common red flags. USAPI’s published specification for HPLC purity is ≥99%. Other tests, when they appear on a lot’s COA, should be read against the specification printed on that report.

HPLC purity

What it measures
The share of the main component relative to related impurities in the sample, expressed as a percentage of the total chromatogram peak area. This is the usual purity figure on a peptide COA.
How it is done
The sample is separated by high-performance liquid chromatography, usually reverse-phase HPLC, and each peak is measured, typically by UV absorbance at a stated wavelength.
What a good result looks like
A numeric purity at or above the stated specification — USAPI’s specification is ≥99% — with the method, wavelength and a chromatogram showing a clear main peak. Impurity peaks should be visible on the baseline if they are present, not cropped out of the figure.
Red flags
No chromatogram; purity given as “pass” or “>99%” with no chromatogram or method; a chromatogram that looks identical to one from another lot; a baseline or impurity region that appears cut off.

Identity by mass spectrometry (LC-MS)

What it measures
Whether the material is the compound on the label, by comparing the measured molecular mass with the expected molecular weight. This is identity testing, not a purity test.
How it is done
The laboratory ionizes the peptide and records its mass-to-charge ratio, often by LC-MS or MALDI-TOF.
What a good result looks like
Both the expected mass and the observed mass are listed, and they agree within the instrument’s stated tolerance. A spectrum or peak list is attached.
Red flags
Only one of the two masses is given; the observed mass does not match the expected mass; no spectrum or peak list; a mass that belongs to a different compound.

Peptide content / net peptide content

What it measures
The share of the vial’s total powder weight that is the peptide itself, excluding counterions (such as TFA or acetate) and residual water. This is different from HPLC purity.
How it is done
Commonly by amino acid analysis or nitrogen (elemental) analysis, and sometimes by a calibrated UV assay.
What a good result looks like
A numeric content — as a percentage, or as milligrams of peptide in the vial — with the method named. A high HPLC purity can still come with a lower net peptide content because salt and water add weight.
Red flags
Content and purity used as if they were the same number; no method; a labeled amount treated as if it were net peptide without a content result when content was supposed to be reported.

Water content (Karl Fischer)

What it measures
Residual moisture in a lyophilized material.
How it is done
Karl Fischer titration, a chemical method that specifically quantifies water.
What a good result looks like
A numeric water content with units (typically % w/w) and a stated specification. Lyophilized peptides are supplied as dry solids, so the value should be low and within the specification on that COA.
Red flags
No water result on a lyophilized lot when the report implies a moisture analysis; a result above the stated specification; “dry” with no number.

Endotoxin (LAL)

What it measures
The level of bacterial endotoxin, a component of the outer membrane of certain bacteria, reported in endotoxin units (EU), often as EU per milligram or per vial.
How it is done
The Limulus amebocyte lysate (LAL) assay, or a recombinant equivalent, detects endotoxin through a clotting or chromogenic response.
What a good result looks like
A numeric result with units, compared with a stated specification, and the method named (for example, gel-clot or kinetic chromogenic LAL).
Red flags
“Pass” or “endotoxin free” with no number or units; no specification; confusing endotoxin testing with sterility testing — they are different tests.

Sterility

What it measures
Whether viable microorganisms are detected in the tested sample under the conditions of the test.
How it is done
The sample is introduced into nutrient media and incubated for a defined period; growth indicates a failing result.
What a good result looks like
When sterility testing is included, the report should state the method, the laboratory, and a clear result such as “no growth” or “meets specification.” Many research-material COAs do not include a sterility test; absence of the test is not the same as a passing result. A sterility result, when present, is a laboratory test result. It does not make a research chemical a drug, and it does not mean the material is for human or veterinary use.
Red flags
The word “sterile” on a label or page without a sterility test on the COA; a pass with no method or laboratory; treating a passing endotoxin or bioburden result as sterility.

Heavy metals / elemental impurities

What it measures
Trace elemental impurities such as lead, arsenic, cadmium and mercury, and sometimes other process-related metals.
How it is done
Typically inductively coupled plasma mass spectrometry (ICP-MS) or a related elemental method. Older reports may use a non-specific “heavy metals” colorimetric test.
What a good result looks like
Named elements, each with a result, units and a limit, plus the method. A complete report lists specific elements rather than a single combined “heavy metals” number.
Red flags
A single vague “heavy metals: pass” with no elements, units or method; a non-specific colorimetric test presented as if it identified each element.

Residual solvents

What it measures
Trace organic solvents remaining from synthesis or purification.
How it is done
Usually gas chromatography, with results compared to stated limits for each solvent.
What a good result looks like
Each solvent is named, with a result and a limit, and the method is stated. “None detected” is meaningful when the detection limit is also given.
Red flags
“Conforms” or “meets limits” with no solvent names or numbers; no residual-solvent data on a report that claims a full impurity panel.

TFA / counterion content

What it measures
The amount of counterion — often trifluoroacetate (TFA) salt or acetate — that balances the peptide’s charge in its salt form. The counterion adds to total vial weight but is not part of the sequence.
How it is done
Commonly ion chromatography, and sometimes 19F NMR for TFA or a dedicated HPLC method.
What a good result looks like
The salt form is identified (TFA or acetate), and the counterion is quantified or a specification is given. This number helps interpret net peptide content.
Red flags
Salt form not identified; a TFA-purified peptide with no mention of TFA; counterion confused with an impurity in the HPLC purity figure.

Appearance

What it measures
The visual description of the material as received by the laboratory, usually a lyophilized cake or powder.
How it is done
Visual inspection under stated conditions, recorded as a short description (for example, white to off-white lyophilized solid).
What a good result looks like
A description that matches the type of material (a dry solid or cake for a lyophilized peptide) and notes any unexpected color or foreign matter. Cake structure can vary from a solid puck to loose powder without, on its own, indicating a quality difference.
Red flags
No appearance line; a description that does not match a lyophilized solid (for example, a liquid when a dry cake is expected); obvious contamination not addressed.

Solubility

What it measures
Whether the lot dissolved in a named laboratory solvent under the conditions the laboratory used.
How it is done
A measured amount is mixed with a stated solvent and the appearance of the solution is recorded.
What a good result looks like
The solvent, conditions and observation are all stated. This is a laboratory observation, not a use or preparation instruction.
Red flags
“Soluble” with no solvent or conditions; treating a solubility line as a protocol for use.

Bioburden / microbial limits

What it measures
The number of viable microorganisms in the sample, often as total aerobic microbial count and yeast and mold count, compared with stated limits. This is a count, not a sterility test.
How it is done
The sample is cultured on appropriate media and colonies are counted after incubation, typically as colony-forming units (CFU) per unit tested.
What a good result looks like
Numeric counts with units, stated limits, and the method named.
Red flags
“Microbial: pass” with no counts or method; a bioburden result presented as a sterility test.

Amino acid analysis

What it measures
The amino acid composition of the peptide after it is broken down into amino acid residues. It can support identity (the residue ratios should match the sequence) and is also used to calculate net peptide content.
How it is done
The peptide is hydrolyzed, then the free amino acids are separated and quantified, typically by chromatography.
What a good result looks like
A table of found versus theoretical residues, with a stated method, and a clear comparison to the expected sequence. Some residues (notably tryptophan and cysteine) can degrade during hydrolysis; a complete report notes how those residues were handled.
Red flags
No theoretical comparison; a composition that does not match the labeled sequence without comment; amino acid analysis used as the only identity test when a mass spectrum is missing and the composition is incomplete.

Other results you may see

Some reports include additional observations. A pH line records the pH of a laboratory solution prepared for the test; it is an observation, not a preparation instruction. Optical rotation, when present, is another identity-supporting measurement. Impurity peaks may be identified by LC-MS (for example, deletion or truncated sequences from solid-phase peptide synthesis). If the salt form is acetate, residual acetic acid may be reported alongside, or instead of, TFA. Treat these the same way as the tests above: look for a method, a specification and a result with units.

Red-flag checklist

Use this list when you open a COA. Any one of these is a reason to ask for a complete, lot-specific report.

  • No lot number, or the lot number does not match the vial label.
  • No laboratory name, or the laboratory cannot be identified.
  • Generic or reused COA: identical chromatograms or identical results across different lots.
  • Purity claimed without a chromatogram.
  • Observed mass does not match the expected mass, or only one of the two is given.
  • No dates, or no signature or approval.
  • No specification, only a bare “pass,” or no method stated.
  • HPLC purity treated as if it were net peptide content.

This guide is general reference for laboratory research. It is not a certificate of analysis and is not a substitute for the lot-specific report. Related pages: COA / Lab Results and the Glossary of research terms. All USAPI products are for research use only and are not for human or veterinary use.

Download: How to Read a Peptide COA

A printable PDF of this guide: every test on a certificate of analysis and what to check. Enter your email and we’ll show the download link and send you a copy. We’ll also send occasional research-library and COA updates; unsubscribe anytime.